This site uses cookies so that we can provide you with the best possible user experience. The information in the cookies is stored in your browser and performs functions such as recognizing you when you return to our site or helping our team understand what sections of the site you find most interesting and useful.
Bibliography
Journal of Oral & Maxillofacial Surgery; 67; 195-199; 2009. Boronat-Lopez A, Carrillo C, Peñarrocha M, Peñarrocha-Diago M. Valencia University Medical and Dental School, Valencia, Spain.
Introduction:
The primary stability of dental implants is related to the bone in contact with the latter, and can be evaluated by resonance frequency analysis.
Material and Methods:
Measurements were made in 133 implants (62 in the upper jaw and 71 in the mandible) of resonance frequency and insertion force to determine implant stability on the day of surgery, with an evaluation of its relationship to different variables.
Results:
The stability quotient of the implants on the day of surgery was 62.1, with an insertion force of 35.7 N. The insertion force was proportional to the resonance frequency, with an increasing stability quotient with growing insertion force. The stability quotient was greater in the larger diameter implants, shorter implants, in mandibular placement and in areas of more compact bone.
Conclusions:
The stability quotient on the day of implant placement is greater in higher bone density areas.